Care and use manual – Waters XBridge Protein BEH, C4, 300A, 3.5 µm Columns User Manual
Page 8
 
8
[ CARE AND USE MANUAL ]
XBridge Protein BEH C
4
, 300Å, 3.5, 5, 10 �m
reduces the possibility of incomplete elution of the protein from 
the column. Second, memory effects may be more pronounced with 
steep gradients. Keep the gradient slope at 1% per column volume 
or less. Third, memory effects may be exacerbated by high flow 
rates. Reduce the flow rate by one half while doubling the gradient 
time to maintain a constant slope. Fourth, memory effects may be 
reduced by changing the organic solvent to incorporate propanol, 
typically 70% propanol:30% acetonitrile as strong solvent. 
Fifth, carryover may be reduced in routine assays with a 
regeneration step including a series of fast gradients from 0–100% 
acetonitrile. The gradients can be as short as 2 column volumes and 
3–5 repetitions may be effective. This “sawtooth” gradient may be 
appended to each injection. Finally, apparent memory effects may 
actually reflect the solubility of the protein in the mobile phase. 
Reducing the amount injected may eliminate the effect.
5. Recovery is often improved by elevating the column
temperature.
Note: Useful general information on column troubleshooting problems may 
be found in HPLC Columns Theory, Technology and Practice, U.D. Neue, 
(Wiley-VCH, 1997), the Waters HPLC Troubleshooting Guide (Literature code 
# 720000181EN) or visit the Waters Corporation website for information on 
seminars (www.waters.com).
VI. COLUMN CLEANING, REGENERATION, 
 
AND STORAGE
a. Cleaning and Regeneration
Changes in peak shape, peak splitting, shoulders on the peak, 
shifts in retention, change in resolution, carryover, ghost peaks, 
or increasing backpressure may indicate contamination of the 
column. Choose a cleaning option that may be expected to 
dissolve the suspected contaminant.
1. All cleaning procedures will be more effective at higher
temperatures. The BEH300 C
4
can be routinely operated at
temperatures as high as 90 °C so it is reasonable to conduct 
cleaning at 70–90 °C.
2. It may be useful to conduct cleaning procedures at one-half
the flow rate typical used with that column. In this way the 
possibility of high pressure events is reduced.
3. The first and simplest cleaning procedure is to run a series of
fast gradients from 0–100% acetonitrile. The gradients can 
be as short as 2 column volumes and 3–5 repetitions may be 
effective. This “sawtooth” gradient may be appended to each 
injection to stabilize routine assays.
4. Several different cleaning solutions may be injected to
strip strongly adsorbed material or particulates from the 
column. Make the largest injection possible with the system 
configuration. With such strong cleaning solutions, it is best to 
disconnect the detector from the column and to direct the flow 
to waste.
a. An injection of 1% formic acid.
b. An injection of 10% formic acid.
c. An injection of either 4M urea or 6M guanidine-HCl.
d. If contamination with lipids is suspected, a strong 
 cleaning option is an injection of tetrahydrofuran. 
5. Flow reversal or backflushing is often suggested as part of a
cleaning procedure. This should be reserved as a last resort. 
It may further damage the column or provide a short-lived 
improvement in performance.
b. Storage
For short-term storage, the column should be stored in the mobile 
phase with 20–50% acetonitrile. For periods longer than four 
days at room temperature, store the column in 100% acetonitrile. 
Immediately after use with elevated temperatures and/or at pH 
extremes, store in 100% acetonitrile for the best column lifetime. 
Do not store columns in highly aqueous (<20% organic) mobile 
phases, as this may promote bacterial growth. If the mobile 
phase contained a buffer salt, flush the column with 10 column 
volumes of HPLC grade water (see Table 1 for common column 
volumes) and replace with 100% acetonitrile for storage. Failure 
to perform this intermediate step could result in precipitation of 
the buffer salt in the column or system when 100% acetonitrile 
is introduced. Completely seal column to avoid evaporation and 
drying out of the bed. 
V. CONNECTING T HE COLUMN TO T HE HPLC
a. Column Connectors and System Tubing Considerations
Tools needed:
§
3/8-inch wrench
§
5/16-inch wrench
Handle the column with care. Do not drop or hit the column on a 
hard surface as it may disturb the bed and affect its performance.