Care and use manual – Waters PoraPak RXN Cartridges User Manual
Page 2
 
[ Care and Use ManUal ]
PoraPak
™
RXN Cartridges and Bulk Care & Use
2
I. Common ApplICAtIons for CleAnup
Bases
Acids
Neutrals
Catch and Elute
Pass Through
Adsorption
PoraPak Rxn CX
PoraPak Rxn CX
PoraPak Rxn RP
TFA removal
Convert from high boiling
point solvent for bases
Catch and Elute
Catch and Elute
PoraPak Rxn CX
PoraPak Rxn CX
Fractionate sample
before Prep for bases
Remove water from post
column fraction for bases
Catch and Elute
Catch and Elute
PoraPak Rxn CX
PoraPak Rxn CX
Technique
Recommended Product
Reaction Products
Reaction Components
Pre-and-Post
Preparative Column
Catch and Elute
In	the	“catch	and	elute”	technique,	the	reaction	mixture	is	loaded	onto	the	cartridge;	the	analytes	of	interest	are	retained	by	the	sorbent.		A	wash	step	follows	to	remove	
additional	reaction	components	from	the	cartridge.	A	strong	solvent	is	used	to	elute	the	analytes	from	the	cartridge.	Sample	concentration	results	when	the	final	elution	
volume is smaller than the load volume.
Pass Through
In	the	pass-through	technique,	the	sorbent	does	not	retain	the	compounds	of	interest,	but	does	retain	some	class(es)	of	compounds	and	impurities,	that	can	be	discarded	
with the column. Fractionation is achieved by the interaction between the impurities in the reaction mixture and the sorbent. No sample enrichment occurs during the 
solid-phase	extraction	(SPE)	step.
Adsorption
In	the	adsorption	chromatography	technique	on	a	reversed-phase	sorbent,	the	neutral	compounds	are	moving	down	the	chromatographic	column	in	the	presence	of	
organic	solvent,	albeit	slowly,	for	small-percent	organic	washes.	Load	the	reaction	mixture	in	high	water	content
(75%	water	for	example),	wash	to	remove	polar	compounds,	and	then	increase	the	organic	content	in	large	increments	to	elute	neutrals.		The	increase	is	determined	by	
the hydrophobicity and MW of the compounds.