[ Care and Use ManUal ]

Waters Sep-Pak DNPH-Silica Cartridge

10

5. Analyze all three cartridges by HPLC.

6. Subtract the value from the blank cartridge from the values

determined from the other two cartridges.

7. Calculate and sum of the total captured DNPH-derivative from

both cartridges 1 and 2.

8. Divide the amount of DNPH-derivative determined from the first

cartridge by the total amount determined form cartridges 1 and
2. Multiply by 100. This is the percentage of DNPH-derivatives
captured on the first cartridge. This value should exceed 95%
otherwise, some of the sample broke through to the second
cartridge.

Example: Measuring Sample Breakthrough

To measure breakthrough:

Expected concentration of formaldehyde is 500 ppbv. Flow rate is
2.0 L/min. A sample volume if 100 liters yields:

Analyte ppmv Carbonyl Collected
concentration x molecular x air volume = µg Carbonyl

weight

Analyte molar volume at 1 atm/25 °C

This calculates to:

0.66 µL x 30.03 g/mole x 100 L = 81 µg formaldehyde

24.46 L/mole

The actual results are shown in Table 7. To calculate the percent
captured on the first sampler, divide the quantity captured on sampler
1 by the total quantity captured, then multiply by 100. Since this
value is less than 95%, and the total carbonyl amount exceeded
2.3 µmoles, breakthrough occurred.

Table 7: Breakthrough Example HPLC Results

In the above example, only a single carbonyl source was present.
Under many test conditions more than one carbonyl source may be
present in significant concentrations. These other compounds will
consume DNPH, effectively reducing the capacity of the sampler for
the compound of interest. To assure that the capacity of the sampler
has not been exceeded, compare the DNPH peak areas of the sample to
a similarly eluted blank. The DNPH peak area in all samples must be
no less than 50% of the DNPH peak area of the blank. This ensures
the sampler capacity has not been exceeded.

Sampler

Amount (µg)

Quantity Captured

Sampler – blank (µg)

Percent Captured on

Sampler

Sampler 1

75.06

75.00

91.8

Sampler 2

6.72

6.66

8.2

Blank