Care and use manual – Waters nanoACQUITY UPLC Columns User Manual
Page 3
 
[ Care and Use ManUal ]
nanoACQUITY UPLC Column Care and Use
3
Carefully grab the teflon tubing with one hand and the BARE FUSED 
SILICA capillary, not the PEEK tubing, with the other hand, to remove 
the Teflon sleeve. Again, be careful NOT to put unnecessary force 
when removing the teflon tubing which could cause the fragile joint 
between transfer tube and packed capillary to separate.
B: CONNECTION
Connecting and proper tightening of 75, 100, or 150 µm,
nanoACQUITY columns to 10 K UPLC system
Flow mobile phase only in the direction indicated by the arrow on the
column label. Connect the column with the direction of the flow arrow 
on the label pointing to the detector or Mass Spectrometer.
1. For a column never installed on a nanoACQUITY system 
  Turn nut till snug then an additional 1/2 turn
2. For a previously installed and removed column from 
 a nanoACQUITY system 
  Hand tight plus 1/8 turn
“Hand Tight” is when the ferrule bottoms in the M-detail fitting. The 
ferrule is pre-staked higher than necessary on the tube, so the ferrule 
must be seated by hand (hand tight) or lightly with a wrench, then 
an additional turn. Note: In virtually every case, gentle wrenching is 
required to get to the “hand tight” state before the final half turn. 
Taking “hand tight” or “finger tight” too literally can result in under 
tightening/leaks. 
Note: DO NOT over tighten, the leak rate will increase and the ferrule 
may get stuck in the port
IV. PreParIng and ConneCtIng 
300 µm nanoaCQuIt Y CaPIllarY Columns
• Carefully remove column from box
•
Connect	the	column	inlet	to	the	injector	utilizing	a	Valco	style 
compression screw
•
Connect	the	column	outlet	to	the	detector	or	Mass	Spectrometer 
using a Valco style compression screw.
V. Care when stoPPIng flow to Column
•
Do	not	stop	the	flow	to	your	nanoACQUITY	column	suddenly.	It	is	
critical that the flow be slowly lowered via the nanoACQUITY system 
controller to prevent column damage.
VI. how to dIagnose and address abnormallY 
hIgh baCkPressure and leaks 
The inlet or the head of a chromatography column experiences the
largest amount of pressure in a LC system. The nanoACQUITY column 
inlet is designed to operate at pressures up to 10 K psi.
The fluidic pressure drops across the column at the outlet of the column
should only experience atmospheric pressure. The outlet connection 
which connects the column to the transfer capillary is designed to 
handle a maximum pressure of 800 psi.
With a higher than normal system pressure, a small leak might be
detected at the outlet of a nanoACQUITY column (see below).
Leak at column outlet
Column
Union
Transfer tubing
Ferrule
Nut
Wrong
Correct