About attached goods outside the standard

attention :

Request the following options when you place an order of Thermo Elite. After

installation the order for these options cannot be accepted.

5. Digital printer OBH60

OBH 60

„ The following information can be printed out by using Digital Printer OBH60.

Character Record Mode

Measured Temperature

Set point

Wait state

Strength of pump

Alarming state

Accumulated time

Time of the record

Graphic Record Mode

The measured temperature of the liquid in the reservoir is recorded by the time based graph.

Immediate Date Record Mode

The information at desired time can be recorded. The contents of the information is the same as
that of Character Record Mode.

„ When the main body of the equipment detects abnormality, information when generated abnormally

is printed.

„ When the power supply is intercepted due to the power failure etc. , the date, the time of the

intercepted point, and the date and the time of the returning point are printed immediately after the
power supply return.

6. Independent temperature overheating prevention device

(BH500 is equipped normally)

OBH 50

„ The equipment is independent safety device besides the one built into the controller. Double safety

measures can be considered.

Range of temperature setting

Range of temperature setting

Range of temperature setting

Range of temperature setting : 0∼399℃ 3-digit digital switch
Sensor

Sensor

Sensor

Sensor : K-thermocouple (Control sensor : Double sensor stored in the same protection tube as

Pt100

Ω

)

to Thermo Elite

to AC100V