110 Enpac Ex Parameters
Memory
Configures the data collector’s memory.
• select cmd - Exit this parameter without selecting
an option.
• Purge - Marks all the points on the route as taken,
sets the current reading to be the previous reading
for each point, and deletes all the off route data.
• Delete route - Deletes the current route in
memory.
• Delete all routes - Deletes all routes and data in
active memory.
• Initialize - Formats the active memory. Note this
will delete all routes and data in active memory.
Options: select cmd (default)
Purge
Delete route
Delete all routes
Initialize
Addnl Measts
The method for advancing when there is additional
measurements.
• Automatic - The Enpac Ex automatically accepts
the additional measurement without pressing the
READ/OK key.
• Manual - The READ/OK key must be pressed to
accept the additional measurement.
Options: Automatic (default)
Manual
Fill Signature
Determines whether to fill in the space between the
signature line and the X axis on spectrum plots.
• On - Fill the space between the signature line and
the X axis.
• Off - Plot spectra as a single curve.
Options: On
Off (default)
Instrument Configuration Parameters
Parameter Name
Description
Values/Comments