12 Dual Probe Holder

Publication GMSI10-UM028A-EN-E - November 2007

11. Apply a film of grease to the exposed surface of the housing o-ring (on base) if

needed. Place the Dual Probe Holder cover on the base, and hand tighten. Be sure to
turn an additional 1/2 turn to seat the enclosure seal.

Replacing Series 9000 Sensor

To remove 9000 sensor from Dual Probe Holder

1. Remove Dual Probe Holder cover.

2. Disconnect wiring leads.

3. Remove stud or bolt and remove sensor.

To install 9000 sensor

1. Clean housing threads.

2. Install 9000 sensor into the base. Use stud or bolt (1/4-28 threads x 3/8 inch) and

thread locking compound. Socket wrench can be used.

3. Apply a film of grease to the exposed surface of the housing o-ring (on base) if

needed. Place the Dual Probe Holder cover on the base, and hand tighten. Be sure to
turn an additional 1/2 turn to seat the enclosure seal.