Load troubleshooting, Repair, Figure 10 cal cart outline – Bird Technologies SCC8 Series-Manual User Manual
Page 30

Maintenance
16
Load Troubleshooting
Repair
Figure 10 Cal Cart Outline
PROBLEM
POSSIBLE CAUSE
CORRECTION
Blower not
operating
Is the load’s AC power cord
connected to the terminal strip?
Connect the power cord.
Is the Cal Cart’s AC power cord
connected to the AC line?
Connect AC power.
Unit turned off.
Set the line switch to ON.
Fuse burnout.
Replace fuse (See "Replacing
Fuses" on page 18).
Coolant leaking
Loose connections.
Tighten drain plug and all
connections.
Worn or cracked tubing.
Replace defective tubing.
Excessive
reflected power
Resistor’s DC resistance has
changed.
Check DC resistance. See "DC
Resistance" on page 12).
WARNING
To avoid personal injury, disconnect the power cord from the AC line before
performing any maintenance, including fuse replacement.
WARNING
Never attempt to connect or disconnect RF equipment from the
transmission line while RF power is being applied.
Leaking RF energy is a potential health hazard.
40"
(1016 mm)
Drain
Plug
Coolant
Gauge
Coolant
Filler Cap
Power
Meter
22"
(559 mm)
39"
(991 mm)
ON/OFF
Switch
Front
Panel
Power Sensor
RF Connector
Handle