Maintenance
18
1. Disconnect the RF line from the HPCC.
2. Remove the unit’s front panel (see "Removing the Front Panel" on
3. Disconnect the sensor cable from the power sensor.
4. Disconnect the RF cable from the sensor output port.
5. Remove the screws on the sensor mounting bracket and remove the sensor
from the mounting bracket.
6. Put the new sensor in the bracket and screw it into place.
Note:
Make sure the arrow on the side of the sensor points towards
the load, and that the end labeled “SOURCE” points towards the front
of the Cal Cart.
7. Connect the RF cable to the sensor end labeled “LOAD”.
8. Connect the sensor cable.
9. Replace the unit’s front panel.
Replacing Fuses
1. Gently pry the fuse drawer out of the AC module.
Note:
The fuse holder does not detach from the AC module.
2. Install the replacement fuses then close and secure the fuse holder.
Fuse 1 is a spare fuse and fuse 2 is the active fuse.
Figure 12 AC Line Fuse
WARNING
To avoid personal injury, disconnect the power cord from the AC line before
performing any maintenance, including fuse replacement.
CAUTION
Failure to install the properly rated fuse may result in equipment damage
or nuisance failures.
AC Line Voltage
Fuse Rating
115 Vac
T630 mA, 5x20 mm Time Lag Fuse
230 Vac
T315 mA, 5x20 mm Time Lag Fuse
Fuse 1
Fuse 2
Fuse
Drawer
AC Module