Replacing fuses, Figure 12 ac line fuse – Bird Technologies SCC8 Series-Manual User Manual

Page 32

Advertising
background image

Maintenance

18

1. Disconnect the RF line from the HPCC.
2. Remove the unit’s front panel (see "Removing the Front Panel" on

page 17).

3. Disconnect the sensor cable from the power sensor.
4. Disconnect the RF cable from the sensor output port.
5. Remove the screws on the sensor mounting bracket and remove the sensor

from the mounting bracket.

6. Put the new sensor in the bracket and screw it into place.



Note:

Make sure the arrow on the side of the sensor points towards

the load, and that the end labeled “SOURCE” points towards the front

of the Cal Cart.

7. Connect the RF cable to the sensor end labeled “LOAD”.
8. Connect the sensor cable.
9. Replace the unit’s front panel.

Replacing Fuses

1. Gently pry the fuse drawer out of the AC module.



Note:

The fuse holder does not detach from the AC module.

2. Install the replacement fuses then close and secure the fuse holder.

Fuse 1 is a spare fuse and fuse 2 is the active fuse.

Figure 12 AC Line Fuse

WARNING

To avoid personal injury, disconnect the power cord from the AC line before

performing any maintenance, including fuse replacement.

CAUTION

Failure to install the properly rated fuse may result in equipment damage

or nuisance failures.

AC Line Voltage

Fuse Rating

115 Vac

T630 mA, 5x20 mm Time Lag Fuse

230 Vac

T315 mA, 5x20 mm Time Lag Fuse

Fuse 1

Fuse 2

Fuse

Drawer

AC Module