Metrohm 840 PC Control 5.0 / Touch Control User Manual

Page 331

Advertising
background image

6

Appendix

PC Control / Touch Control

319

The titration cell should be filled with solvent before the sample is

added so that the sample can be pipetted directly into the liquid (tip

immersed).
Command sequence:

LQH

change port

MOVE

move to titration cell

LIFT

move to work position

Eject sample
The sample should be ejected at a reduced dosing rate (< 10 mL/min).

The appropriate settings can be made in the liquid handling command

under Rate.
Command sequence:

LQH

End volume

Rinse
In order to keep sample solution carryover as low as possible, rinsing

with solvent can be carried out after the sample has been ejected. This

is particularly recommended for non-aqueous samples.
In preparation for the next pipetting process the cylinder should be

completely ejected.
Command sequence:

LQH

fill or aspirate

LQH

eject