Select a connection or a device and open the corresponding

configuration dialog with [Properties].

The settings for the USB Sample Processor, the tower and the

Swing Head are described on the following pages.
For the MSB connections you can define when the request to carry

out the preparation function (PREP command) for the connected

doing devices is to be displayed. The preparation of the exchange or

dosing unit is described in Section 3.24.2. For the description of the

settings please consult the online help.
The Stirrer type and Serial number are shown for the connected

stirrers.
The Dosing device type and the Serial number (for type 8xx dosing

devices only) are shown for the connected dosing devices.
You get to the overview of all the racks available with [Rack tables]

(see p. 63).
With [Running time] you access the overview of the running time

meter. The running time meter adds together the time while the USB

Sample Processor is "busy" i.e. when a single action is being carried

out. After expiry of the time limit set, a message with the advice for a

service is displayed. Metrohm recommends to carry out a service after

1000 operating hours. A reset of the running time meter can only be

carried out by a Metrohm service technician.