Metrohm 815 Robotic USB Sample Processor XL User Manual
Page 26

3.4 Mounting the Swing Head to the Sample Processor
■■■■■■■■■■■■■■■■■■■■■■
18
■■■■■■■■
815 Robotic USB Sample Processor XL
■
Plug the Mini DIN plug into the socket 'Swing Head'.
Mounting the Swing Head
1
2
3
4
1 Screw the Swing Head tightly to the holder plate of the guide chain.
2 Clamp the Swing Head between the guide jaws.
3 Screw the Swing Head to the guide jaws with the screws previously
loosened.
4 Place the connection cable into the guide chain (see Chapter 3.4.2,
Mounting the Swing Head with tower extension
For certain applications, it may be necessary to mount the 815 Robotic
USB Sample Processor XL at a higher position. A 6.2058.010 tower
extension can be mounted on the lift for this purpose.
1 Fasten the tower extension to the lift guide of the Sample Processor
in accordance with the following illustration and fix it in place with
the screws provided.