3 Installation
815 Robotic USB Sample Processor XL
■■■■■■■■
21
3.4.3
Mounting the robotic arm reinforcement
In cases in which it is necessary to take up or strip off a filter or a pipetting
tip on the robotic arm, you must mount a 6.2058.090 robotic arm
reinforcement. The reinforcement provides the robotic arm with stable
resistance when picking up or stripping the tools used and prevents it
from bending.
The reinforcement has to be placed before mounting a robotic arm. Pro-
ceed as follows:
2
1
6.2058.090
Figure 12
Mounting the 6.2058.090 robotic arm reinforcement
1 Shift the reinforcement from the front over the holder plate of the
Swing Head.
2 Screw the reinforcement to the lift guide with the screws provided.