6.3 Remote interface

■■■■■■■■■■■■■■■■■■■■■■

50

■■■■■■■■

815 Robotic USB Sample Processor XL

6.3.1

Pin assignment of the remote interface

13

1

14

25

1

13

14

25

Figure 38

Pin assignment of the remote socket and plug

The above presentation of the pin assignment of a Metrohm remote inter-
face applies not only for the remote box, but also for all Metrohm devices
with 25-pin D-Sub remote connection.

Inputs

+5 V

approx. 50 k

Ω Pull-up

tp

t

p

>20 ms

active = low, inactive = high

The input lines can be scanned with the SCAN command.

Outputs

Open Collector

tp

t

p

>200 ms

active = low, inactive = high

I

C

= 20 mA, V

CEO

= 40 V

+5 V: maximum load = 20 mA

The output lines can be set with the CONTROL command.

Table 1

Inputs and outputs of the remote interface

Assigment

Pin No.

Assigment

Pin No.

Input 0

21

Output 0

5

Input 1

9

Output 1

18

Input 2

22

Output 2

4

Input 3

10

Output 3

17

Input 4

23

Output 4

3

Input 5

11

Output 5

16