10 equipping the titration head, Figure 21 – Metrohm 815 Robotic USB Sample Processor XL User Manual
Page 39
Advertising

■■■■■■■■■■■■■■■■■■■■■■
3 Installation
815 Robotic USB Sample Processor XL
■■■■■■■■
31
3.10
Equipping the titration head
Mounting the aspiration and rinsing tubings
1
2
3
6.1805.060
6.1805.510
6.1543.170
Figure 21
Installing the rinsing tubings and the aspiration tip
Proceed as follows:
1 Connect the rinsing nozzles
■
Connect the three rinsing tubings that are already connected to
the Tower 2 distributor to the rinsing nozzles already mounted on
the titration head.
2 Insert the aspiration tip
■
Insert the 6.1543.170 aspiration tip into the opening left on
the front of the titration head.
3 Connect the aspiration tubing
■
Connect the 6.1805.510 aspiration tubing already connected to
the distributor with the aspiration tip.