2 Installation

2.2 Mains

connection

2.2.1

Mains voltage and fuses

The 830 Interface has a power supply which automatically adjusts itself

to the existing mains voltage (100

…

240 V) and frequency (50

…

60 Hz).

It is equipped with an electronic overload protection device and also

has two fuses; however, these should only be exchanged by Metrohm

service technicians.

2.2.2 Mains

cable

The instrument is supplied with one of three mains cables

•

6.2122.020 with plug SEV 12 (Switzerland,

…

)

•

6.2122.040 with plug CEE(7), VII (Germany,

…

)

•

6.2133.070 with plug NEMA 5-15 (USA,

…

)

which are three-cored and fitted with a plug with an earthing pin. If a dif-

ferent plug has to be fitted, the yellow/green lead (IEC standard) must

be connected to the earthing socket 2 (protection class 1).

Any break in the earthing inside or outside the instrument can make it

a hazard!

Plug the mains cable into mains connection plug 5 at the 830 IC Inter-

face (see Fig. 3).

2.2.3

Switching the instrument on/off

The 830 IC Interface is switched on and off using mains switch 3 (see

Fig. 3). When the instrument is switched on a start-up check routine is

running the mains pilot lamp 1 flashes twice slowly and remains on.

830 IC Interface / 8.830.1003 Instructions for Use

8