7 833 ic liquid handling ultra filtration unit, 833 ic liquid handling ultra filtration unit, Fig. 10 – Metrohm 830 IC Interface User Manual
Page 23: Connection of 819, 820, 818, 833 (suppressor unit), And 833 (ultra filtration unit)

2.4 Connection of external instruments
2.4.7
833 IC Liquid Handling Ultra Filtration Unit
Example
Analysis of anions on a modular system with chemical suppression af-
ter preceding sample filtration, for strongly contaminated samples.
Instruments
•
2.830.0020 830 IC Interface
•
2.819.0110 819 IC Detector
•
2.820.0210 820 IC Separation Center 820 with 1 injector and
column
heater
•
2.818.0110 818 IC Pump
•
2.833.0020 833 IC Liquid Handling Suppressor Unit
•
2.833.0050 833 IC Liquid Handling Ultra Filtration Unit
Interconnection
833 Ultra Filtration
833 Suppressor
819
820
818
830
PC
6.2134.100
6.2134.130
6.2115.070
6.2134.090
6.2125.090
6.2143.220
6.2134.040
Fig. 10:
Connection of 819, 820, 818, 833 (Suppressor Unit) and
833 (Ultra Filtration Unit)
830 IC Interface / 8.830.1003 Instructions for Use
17