4 833 ic liquid handling suppressor-unit, 833 ic liquid handling suppressor-unit, Fig. 7 – Metrohm 830 IC Interface User Manual
Page 20: Connection of 2, 818 and 833 (suppressor unit) (mic 3)

2 Installation
2.4.4
833 IC Liquid Handling Suppressor-Unit
Example
Simultaneous analysis of cations and anions on a modular system with
two independent detectors.
Instruments
The setup of the instruments corresponds to the Modular IC System 3
(MIC 3 Advanced).
•
2.830.0020 830 IC Interface
•
2.819.0010 819 IC Detector for System 1 (cations)
•
2.819.0110 819IC Detector für System 2 (anions)
•
2.820.0220 820 IC Separation Center with 2 injectors and
column
heater
•
2.818.0110 818 IC Pump, for System 1 (cations)
•
2.818.0110 818 IC Pumpe, for System 2 (anions)
•
2.833.0020 833 IC Liquid Handling Suppressor Unit,
for System 2 (anions)
Interconnection
819-2
819-1
820
818-1
830
PC
6.2134.100
6.2115.070
6.2134.090
6.2125.090
6.2143.200
833
818-2
6.2125.090
6.2115.070
6.2134.090
6.2128.180
6.2134.130
6.2134.130
6.2134.040
Fig. 7:
Connection of 2
Ч
819, 820, 2
Ч
818 and 833 (Suppressor Unit)
(MIC 3)
830 IC Interface / 8.830.1003 Instructions for Use
14