3.2 Settings in the "830 IC Interface" window
3.2
Settings in the "830 IC Interface" window
By clicking the 830 icon in the toolbar with the left mouse button or by
clicking this icon with the right mouse button and selecting the
Open
menu item the
830 IC Interface
window for parameter settings is opened.
It consists of three tabs
Init
,
RS settings
, and
Links
.
Init
The
Init
tab of the
830 IC Interface
window contains data acquisition pa-
rameters for the 830 IC Interface.
Label
Optional label to name the interface with
maximum 8 characters.
ROM version
Instrument program version number .
Measuring per second
Number of data points measured per s.
Entry range:
10, 20, 30, 50, 60 points/s
Range
Range for AD converter .
Example: converts
±
2500 mV to 2
24
bits
Entry range:
2500, 1250, 625, 312.5, 156.25, 78.125,
39.062 mV
<SET>
Send current parameters to 830 IC Interface.