Metrohm 830 IC Interface User Manual

Page 17

Advertising
background image

2.4 Connection of external instruments

>CONFIG/RS settings
RS control: on

•

709 IC Pump:

Switch on external control of the 709 IC Pump using the

[ EXT. ] key.

•

752 IC Pump Unit, 753 IC Suppressor Module,

754 IC Dialysis Unit:

Switch on external control via remote interface using the

[ Remote ] switch.

•

817 Bioscan:

This instrument is already pre-configured, no additional set-

tings must be made.

•

761 Compact IC:

This instrument is already pre-configured, no additional set-

tings must be made.

•

750 IC Autosampler:

Always switch on the 830 IC Interface first and then the 750

Autosampler.

•

766 IC Sample Processor:

>RS232 settings
handshake: SWchar

>RS232 settings
RS control: on

•

Triathlon Autosampler:

This instrument is already pre-configured, no additional set-

tings must be made.

•

812 IC Valve Unit:

This instrument is already pre-configured, no additional set-

tings must be made.

•

816 IC Eluent Selector:

Switch on external control via RS232 interface using the

[ EXT ] switch..

•

828 IC Dual Suppresso:

Switch on external control via remote interface using the

[ Remote ] switch.

•

837 IC Degasser:

This instrument is already pre-configured, no additional set-

tings must be made.

5 Create new system in «IC Net»

•

Start the «IC Net» program.

•

Create a new system file with the selected instruments (de-

tails see IC Net Instructions for Use).


830 IC Interface / 8.830.1003 Instructions for Use