6 connecting the bottle with the absorber solution, Connecting the bottle with the absorber solution – Metrohm 920 Absorber Module User Manual

Page 60

Advertising
background image

4.4 Dosino for the absorber solution

■■■■■■■■■■■■■■■■■■■■■■

52

■■■■■■■■

920 Absorber Module

3 Connecting the column's outlet

■

Screw the PTFE capillary (6.1803.030) with the yellow colored
sleeve (fed in from the right side of the instrument) in place on
the trap column's outlet using a pressure screw (6.2744.010).

Inserting the trap column in the instrument

Insert the Metrosep A Trap 1 – 100/4.0 in the instrument as follows:

1 Fasten the Metrosep A Trap 1 – 100/4.0 in place in the front holding

bracket (4-4) of the 920 Absorber Module using the column head.

Next step

Connect the loose end of the PTFE capillary, 0.5 mm ID / 3 m
(6.1803.030) to the Combustion Module's outlet (see Chapter 4.10.2,
page 76)
.

4.4.6

Connecting the bottle with the absorber solution

The bottle with the absorber solution is connected to port 2 of Dosinos 1.

Accessories

For this step you need:

■

Clear glass bottle 2 L (6.1608.070) filled with the absorber solution

■

Bottle attachment (6.1602.160)

■

FEP tubing M6 for 920, 150 cm (6.1838.020) fastened to port 2 of the
Dosino for the absorber solution.

■

Adsorber tube (6.1609.000)

■

Cotton

■

Adsorber granulate