2 connecting the bottle with the standard solution, Connecting the bottle with the standard solution – Metrohm 920 Absorber Module User Manual

Page 66

Advertising
background image

4.5 10-port valve

■■■■■■■■■■■■■■■■■■■■■■

58

■■■■■■■■

920 Absorber Module

Connect
capillary/tub-
ing at port
…

… to

10

Waste container

4.5.2

Connecting the bottle with the standard solution

The bottle with the standard solution is connected to the 10-port valve.

Accessories

For this step you need:

■

Clear glass bottle 100 mL (6.1608.050) filled with standard solution

■

Bottle attachment (6.1602.150), contains 3 pressure screws

Connecting the bottle with the standard solution

6.1608.050

6.1602.150

6.1602.150

6.1803.160

1 Connecting the aspirating capillary for the standard solution

■

Slide a pressure screw over the loose end of the capillary connec-
ted to port 2 of the 10-port valve (13-3).

■

Thread the capillary through one of the three openings on the
bottle attachment.

■

Screw the bottle attachment onto the bottle.

■

Push the capillary far enough into the bottle that it reaches the
bottom of the bottle. Then fix it in place with a pressure screw.