1 connecting the ports, Connecting the ports, Figure 17 – Metrohm 920 Absorber Module User Manual
Page 70: Connecting the 6-port valve
4.6 6-port valve
■■■■■■■■■■■■■■■■■■■■■■
62
■■■■■■■■
920 Absorber Module
Figure 17
Connecting the 6-port valve
1
PTFE capillary, 0.75 mm ID / 3 m
(6.1803.160)
Connection between the 6-port valve and
the waste container.
2
PEEK capillary, 0.5 mm ID / 15 cm
(6.1831.040)
Connection between the 6-port valve and
the 10-port valve.
17 PTFE capillary, 0.5 mm ID / 3 m
(6.1803.030)
Approx. 40 cm. Connection between the
Metrosep I Trap 1 - 100/4.0 trap column
(6.1014.200) and the 6-port valve.
18 PTFE capillary, 0.5 mm ID / 3 m
(6.1803.030)
Connection between the 6-port valve and
the 930 Compact IC Flex.
4.6.1
Connecting the ports
Connectors
The ports of the 6-port valve are connected as follows:
Port …
… is used for …
… is connected to
1
Inlet for sample, standard solution and check
standard
10-port valve
2
Outlet for waste
Waste container
3
Sample loop
6-port valve – Port 6
4
Outlet for sample, standard solution and check
standard
Injection valve in the IC instrument