1 connecting the ports, Connecting the ports, Figure 17 – Metrohm 920 Absorber Module User Manual

Page 70: Connecting the 6-port valve

Advertising
background image

4.6 6-port valve

■■■■■■■■■■■■■■■■■■■■■■

62

■■■■■■■■

920 Absorber Module

Figure 17

Connecting the 6-port valve

1

PTFE capillary, 0.75 mm ID / 3 m
(6.1803.160)
Connection between the 6-port valve and
the waste container.

2

PEEK capillary, 0.5 mm ID / 15 cm
(6.1831.040)
Connection between the 6-port valve and
the 10-port valve.

17 PTFE capillary, 0.5 mm ID / 3 m

(6.1803.030)
Approx. 40 cm. Connection between the
Metrosep I Trap 1 - 100/4.0 trap column
(6.1014.200) and the 6-port valve.

18 PTFE capillary, 0.5 mm ID / 3 m

(6.1803.030)
Connection between the 6-port valve and
the 930 Compact IC Flex.

4.6.1

Connecting the ports

Connectors

The ports of the 6-port valve are connected as follows:

Port …

… is used for …

… is connected to

1

Inlet for sample, standard solution and check
standard

10-port valve

2

Outlet for waste

Waste container

3

Sample loop

6-port valve – Port 6

4

Outlet for sample, standard solution and check
standard

Injection valve in the IC instrument

Advertising