Figure 15, Deaerate the high pressure pump – Metrohm 882 Compact IC plus – Cation User Manual
Page 36

3.9 High pressure pump
■■■■■■■■■■■■■■■■■■■■■■
28
■■■■■■■■
882 Compact IC plus – Cation
5
1
2
6
7
5
3
4
5
Figure 15
Deaerate the high pressure pump
1
Syringe 10 mL (6.2816.020)
For aspirating the eluent.
2
Luer connector
Part of the purging needle (6.2816.040)
3
Purging needle (6.2816.040)
4
Deaerating capillary
5
PEEK pressure screws, short
(6.2744.070)
6
Purge valve
7
Purge valve rotary knob
1 Connecting the purging needle
■
Push the end of the purging needle (15-3) over the end of the
deaerating capillary (15-4) on the purge valve.
2 Connecting the syringe
■
Insert syringe (15-1) in the Luer connector (15-2) of the purging
needle (see Figure 15, page 28).
3 Opening purge valve
■
Open the rotary knob (15-7) by approx.
½ rotation counterclock-
wise.
4 Setting the flow rate
■
Start MagIC Net™ (if not yet started).
■
Ensure that the eluent aspiration tubing is immersed sufficiently in
the eluent.
■
Let the high pressure pump run.