6.1 Problems and their solutions
■■■■■■■■■■■■■■■■■■■■■■
64
■■■■■■■■
882 Compact IC plus – Cation
Problem
Cause
Remedy
Separation column – dead
volume at column head.
■
Install the separation column in reverse
flow direction (if permitted by the leaflet)
and rinse into a beaker.
■
Replace separation column (see "Connect-
ing and rinsing the separation column",
page 38).
Precision problems -
significant scatter-
ing of the measured
values
Injection valve – sample
loop.
Check installation of the sample loop (see
Chapter 3.12.1, page 31).
Sample – rinsing volume
too low.
Increase rinsing time (see Chapter 5.9, page
58).
Injection valve – defective.
Request Metrohm Service.
Unexpected change
to the retention
times in the chroma-
tograms
Separation column –
diminished separating effi-
ciency.
■
Regenerate separation column (see Chap-
ter 5.10.4, page 61).
■
Replace separation column (see "Connect-
ing and rinsing the separation column",
page 38).
High pressure pump –
defective.
Request Metrohm Service.