3.14 Guard column

■■■■■■■■■■■■■■■■■■■■■■

36

■■■■■■■■

882 Compact IC plus – Cation

Note

Metrohm recommends, always to work with guard columns. These pro-
tect the separation column and can be exchanged regularly.

Connecting and rinsing the guard column

1 Connecting the guard column

Caution

When inserting the guard column, always ensure that it is inserted
correctly corresponding to the flow direction (if indicated).

■

Remove sealing caps and/or stoppers from the guard column.

■

Fasten the input of the guard column to the column inlet capillary
using a short PEEK pressure screw (6.2744.070).

■

In case the guard column is mounted on the separation column
with a connecting capillary, connect the included connection
capillary, which is included with the guard column, to the output
of the guard column using the PEEK pressure screw, which is .also
included.

2 Rinsing the guard column

■

Place beaker under the outlet capillary of the guard column.

■

Set the flow rate of the high pressure pump according to the data
given in the leaflet of the separation column.

■

Start the high pressure pump and rinse the guard column approx.
5 minutes with eluent.

■

Switch off the high pressure pump again.