Figure 20, Pump head – removing the piston – Metrohm 882 Compact IC plus – Cation User Manual
Page 55

■■■■■■■■■■■■■■■■■■■■■■
5 Handling and maintenance
882 Compact IC plus – Cation
■■■■■■■■
47
Cleaning/replacing the zirconium oxide piston
Clean one piston after the other as follows:
1 Removing the piston cartridge from the pump head
Loosen the piston cartridge with a wrench and unscrew from the
pump head by hand.
2
1
Figure 20
Pump head – removing the piston
1
Pump head
2
Piston cartridge
2 Dismantling the piston
Caution
On the inside of the piston cartridge there is a taut spring than can
jump out of the piston cartridge if suddenly loosing tension.
When opening the piston cartridge, hold pressure towards the
spring and unscrew carefully.
■
Loosen the screw of the piston cartridge with a wrench and
unscrew carefully by hand and by holding pressure towards the
taut spring.
■
Remove the zirconium oxide piston and lay on a tissue.
■
Remove the spring retainer, spring and the inner plastic sleeve
from the piston cartridge and lay by.