3 Installation

919 IC Autosampler plus

■■■■■■■■

13

This is how you install the needle and the capillary:

1 Removing the pressure screw

Loosen and remove the pressure screw (7-3) screwed onto the nee-
dle holder.

2 Inserting the needle

■

Insert the needle from above partway into the opening of the
needle holder (7-5).

■

Slide the PEEK ferrule (7-4) over the needle from above. The nar-
row side of the seal must face upwards.

3 Fastening the needle

■

Screw the pressure screw (7-3) into the needle holder. Lightly
push the needle upwards from below during the process.

■

Tighten the pressure screw in the needle holder by hand (do not
use tools!).

4 Connecting the capillary

■

Slide the 6.2744.070 PEEK pressure screw (7-2) over the end of
the capillary.

■

Manually tighten the PEEK pressure screw with the capillary to the
pressure screw of the needle holder. The capillary must be pushed
in while doing so.