3 Installation
919 IC Autosampler plus
■■■■■■■■
17
NOTE
Pump tubings are consumables. The lifetime of the pump tubings
depends on the contact pressure, among other factors.
3.5
Filtration cell holder
Figure 10
Installing the filtration cell holder
The filtration cell holder (6.2057.030) can be mounted on the side wall of
the tower, see above.
First remove the second and third screws from the bottom on the side
wall. Then fix the filtration cell holder in place with the two screws sup-
plied.
3.6
Mounting the safety shield
For reasons of safety, it is indispensable that you install the safety shield
(6.2751.110) supplied. A serious risk of injury exists if anyone reaches into
the working area of the instrument.
CAUTION
The 919 IC Autosampler plus may not be operated without a safety
shield.
Use the accompanying hex screws and the hex key to mount the safety
shield according to the following figure.