Figure 9, Inserting the tubing cartridge – Metrohm 919 IC Autosampler plus User Manual
Page 23

■■■■■■■■■■■■■■■■■■■■■■
3 Installation
919 IC Autosampler plus
■■■■■■■■
15
3 Connect the aspiration side
Place a 6.2744.030 tubing olive (8-2) on the aspiration side of the
pump tubing.
4 Connect the pressure side
■
Slide the union nut (8-6) of the 6.2744.160 pump tubing connec-
tion (without filter) onto the pump tubing.
■
Select a suitable adapter (8-7) (depends on the outer diameter of
the pump tubing) and slide it onto the pump tubing (see Table 1,
page 16).
■
Place the tubing olive (8-8) onto the pump tubing.
■
Screw the union nut (8-6) tight on the tubing olive (8-8).
5 Insert the tubing cartridge
■
Hang the tubing cartridge in the mounting bolt and press it
underneath in the cartridge holder until the snap-action lever
snaps in.
Figure 9
Inserting the tubing cartridge