4 installing the peristaltic pump, Installing the peristaltic pump, Figure 8 – Metrohm 919 IC Autosampler plus User Manual
Page 22: Installing the pump tubing

3.4 Installing the peristaltic pump
■■■■■■■■■■■■■■■■■■■■■■
14
■■■■■■■■
919 IC Autosampler plus
3.4
Installing the peristaltic pump
1
2
3
4
5
6
7
8
1
9
10
3
6.2744.160
Figure 8
Installing the pump tubing
1
PEEK pressure screws, short
(6.2744.070)
2
Tubing olive (6.2744.030)
3
Stopper
The colors of the stoppers indicate the inner
diameter of the pump tubing.
4
Tubing cartridge (6.2755.000)
5
Contact pressure lever
6
Union nut
7
Adapter
8
Tubing olive
9
Pump tubing
10 Snap-action lever
Mount the pump tubing as follows:
1 Remove the tubing cartridge
Release the tubing cartridge from the cartridge holder by pressing
the snap-action lever and unhooking it from the mounting bolts
(5-1).
2 Insert the pump tubing
■
Press the contact pressure lever all the way down.
■
Place the pump tubing in the tubing cartridge. The stoppers (8-3)
must snap into the corresponding holder of the tubing cartridge.