6.4.3 Main memory (RAM)

This diagnosis step completes a non-destructive test of the entire area of
the RAM contents (main memory).

• Prepare the unit for diagnosis (see Chap. 6.4.2).

• If necessary, press <Ð> several times until

diagnosis

>RAM test

• <ENTER>

If no errors are found, you will see the following on the display :

>RAM test

RAM test ok

• <ENTER>

6.4.4 Display

This diagnosis step allows you to check the LEDs and the display for oper-
ability.

• Prepare the unit for diagnosis (see Chap. 6.4.2).

• If necessary, press <Ð>several times until

diagnosis

>display test

• <ENTER>

After you press key <ENTER>, the program runs automatically through a test run for optical

checking of the LEDs and the display.
All LEDs blink consecutively for a brief period.
The background lighting of the display is switched off for a short period and back on again.
The power-up test pattern (each pixel active) is displayed.
Both lines of the display are cleared.
Both lines of the display consecutively display characters " # ", "H" and, lastly, "I".
Both lines display the continuous moving text

"0123456789ABCDEFGHIJKLMNOPQRSTUVWXYZ" from right to left.

• The test run can be halted and started again by pressing key <5>.

• The test is quit with key <QUIT> or <STOP>.

diagnosis

>display contrast test

• <ENTER>

After you press the <ENTER> key, the following display is shown, whereby the contrast of the

display is varied constantly between bright and dark.

>display contrast test

774 Oven Sample Proc.

• The test can be quit with key <QUIT> or <STOP>.