1 configuration – Metrohm 774 Oven Sample Processor User Manual
Page 69

5.3 Menu organization
774 Oven Sample Processor, Instructions for use
63
CONFIG
5.3.1 Configuration
Main Menu:
configuration
>auxiliaries
Open submenu with <ENTER>
configuration
>oven settings
Use <
Ï
> or <Ð> to move up or down one menu item
configuration
>rack definitions
Use <HOME> or <END> to move to first or last menu
item respectively
configuration
>dosing units
Return to normal state with <QUIT>
configuration
>RS232 settings
configuration
>auxiliaries
Basic Settings Submenu
Open the submenu with <ENTER>
>auxiliaries
dialog: english
Choice of dialog language
english, deutsch,
français,
español
>auxiliaries
display contrast 3
Setting display contrast
0…3…7
0 = no contrast
7 = large contrast
>auxiliaries
beeper: ON
Turn acoustic warning signal on or off
ON,OFF
>
auxiliaries
device label ********
Instrument label
8 ASCII characters
>auxiliaries
program 5.774.0010
Program version
read
only
The following three entries become effective after a RESET or switching the
changer on again.
>
auxiliaries
max. lift way 90
mm
Max. stroke path for lift
0…90…100 mm
This setting of the max. lift path is important as regards safety. Correctly en-
tering this value can prevent the puncture needle breaking off since the lift
cannot be moved lower than as far as the specified position.
use
<QUIT>
to access next
higher level
use
<QUIT>
to access next
higher level