Metrohm 774 Oven Sample Processor User Manual

Page 72

Advertising
background image

5 Detailed description

774 Oven Sample Processor, Instructions for use

66

>dosing units

1

max. rate 1

160 ml/min

Max. dosing speed (depends on bu-
rette size)

0.01…160 ml/min

>dosing units

1

tube length 1

1000 mm

Length of the tubing on Dosino port 1

0,

1000, 30000 mm

>dosing units

1

tube diameter 1 2 mm

Diameter of the tubing on Dosino port
1

0.1…2.0…20 mm

>dosing units

1

max. rate 2

160 ml/min

Max. dosing speed (dependends on
burette size)

0.01…160 ml/min

>dosing units

1

...........................until Port 4

Enter the tubing diameter for all four
ports of a Dosino

Only the dosing speed is relevant for the 685 Dosimat; the other parameters
are ignored.


Configuration

>RS232 settings

Submenu for the serial interface parameters

Open the submenu with <ENTER>

>RS232 settings

Baud Rate:

9600

Transmission speed in baud

300,600,

1200,2400,

4800,9600

>RS232 settings

data bit:

8

Number of data bits

7,8

>RS232 settings

stop bit:

1

Number of stop bits

1,2

>RS232 settings

parity: none

Parity selection

even, odd, none

>RS232 settings

handshake: HWs

Handshake selection

HWs,HWfull,

SWchar,SWline,

none

>RS232 settings

character set:

IBM

Character set for printer or PC

IBM,HP,Epson,

Seiko,Citizen

use

<QUIT>

to access the

next highest

level

use

<QUIT>

to access the

next highest

level

Advertising