3 Preparing the MME pro for use

6.1246.120 Multi-Mode-Electrode pro

■■■■■■■■

15

■

Mercury

■

Mercury drop catcher (6.2406.000)

1

2

3

4

Figure 5

MME pro – Filling in mercury

1 Placing the MME pro into the electrode stand

■

Place the electrode stand into the drip pan.

■

Place the MME pro into the electrode stand.
The mercury refill opening is pointing upwards.

2 Opening the screw cap

■

Open the screw cap (1-7) at the mercury refill opening with the
wrench and remove it.

3 Filling in mercury

Using the syringe, carefully fill in the mercury into the mercury reser-
voir.

Note

Only fill in mercury until it reaches the making max. fill.