Metrohm Multi-Mode Electrode pro User Manual

Page 27

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Preparing the MME pro for use

6.1246.120 Multi-Mode-Electrode pro

■■■■■■■■

19

HMDE test

1

In the 797 VA Computrace Software, open the Computrace
Control
with Utility

Computrace Control.

Select the electrode type HMDE and switch on the electrode with
[New drop].

At the tip of the capillary, a mercury drop is forming, which serves as
the actual electrode.

2

Watch the drop formation.
Drop formation is functioning properly when the old drop is
knocked off and a new drop is formed. The drop must neither
become larger over time nor fall off.

Advertising