5 Maintenance

6.1246.120 Multi-Mode-Electrode pro

■■■■■■■■

29

Note

If you use another software than 797 VA Computrace, then follow the
instructions on adjusting the needle valve in the manual of your instru-
ment.

For adjusting the needle valve you need the wrench (6.1247.220) or a
coin which will fit the slot of the screw.

1 Preparation

■

Place an empty measuring vessel.

■

Set the inert gas supply to 1.0 to 1.2 bar, in order to pressurize
the electrode.

■

In the 797 VA Computrace software, open the dialog Compu-
trace Control
with Utility

▶ Computrace Control and select

the electrode type HMDE.

■

As soon as a drop of mercury appears at the tip of the capillary,
fill approximately 20 mL of water and a few drops of KCl-solution
into the measuring vessel.

2 Rough adjustment

■

With the wrench, rotate the screw of the needle holder (1-1)
clockwise until the mercury stops dropping.
Check: After each half rotation of the screw, knock off the drop
and check if a new drop is forming.

3 Fine adjustment

■

Rotate the screw slowly counterclockwise until a new drop is
forming.

■

Reseal the capillary with a small rotation in clockwise direction
(ca. 10°).

■

Rotate the screw clockwise by another quarter rotation (90 –
100°).

The MME pro is now ready for operation.