Figure 11, Mme pro – mounting the capillary – Metrohm Multi-Mode Electrode pro User Manual

Page 44

Advertising
background image

5.2 Large maintenance

■■■■■■■■■■■■■■■■■■■■■■

36

■■■■■■■■

6.1246.120 Multi-Mode-Electrode pro

1

2

3

4

Figure 11

MME pro – Mounting the capillary

1 Placing the MME pro into the electrode stand

■

Place the electrode stand into the drip pan.

■

Place the MME pro into the electrode stand.
The thread for the capillary retaining screw is pointing upwards.

2 Inserting the silicone seal

■

Insert the silicone seal. Use the capillary retaining screw to push it
into position.

■

Place the capillary retaining screw loosely into the opening.

Advertising