5.1 Small maintenance

■■■■■■■■■■■■■■■■■■■■■■

28

■■■■■■■■

6.1246.120 Multi-Mode-Electrode pro

4 Replacing the needle

Note

Do not touch the needle with bare hands!

Each needle is individually wrapped.

■

Pull the old needle out of the membrane of the needle holder.

■

Cut open the wrapper of the new needle with the scissors and
unwrap the white plastic part only. Do not yet completely remove
the wrapper.

■

Insert the needle into the hole of the membrane and press in as
far as the white marking. Now, remove the wrapper.

5 Inserting the needle holder

■

Insert the needle holder with the new needle into the MME pro.
Press in as far as it will go.

6 Removing the mounting ring

■

Remove the mounting ring turning it counterclockwise.

7 Fastening the needle holder

■

Use the wrench to tighten the needle holder by turning it clock-
wise until the black sealing ring becomes visible in the electrode
body.

See also the video Replace the needle in the multimedia guide Electrodes
in Voltammetry
.

5.1.5

Adjusting the needle valve

Adjusting the needle valve

Before you can adjust the needle valve, the following requirements must
be fulfilled:

■

The MME pro is mounted onto the measuring head of the instrument
and the inert gas supply is connected.

■

An empty measuring vessel is placed beneath the MME pro.

■

The instrument is connected to a PC on which the 797 VA Computrace
software is installed and running.