5.2 Large maintenance

■■■■■■■■■■■■■■■■■■■■■■

32

■■■■■■■■

6.1246.120 Multi-Mode-Electrode pro

1

2

3

4

Figure 9

MME pro – Removing the capillary

1 Placing the MME pro into the electrode stand

■

Place the electrode stand into the drip pan.

■

Place the MME pro into the electrode stand.
The capillary retaining screw is pointing upwards.

2 Removing the capillary retaining screw

Loosen the capillary retaining screw and remove it.

3 Removing the capillary

Pull capillary and the seal out of the MME pro.

4 Removing the seal

Pull the capillary out of the seal.

The seal can be reused if it is undamaged.