5 Maintenance

6.1246.120 Multi-Mode-Electrode pro

■■■■■■■■

31

■

Drip pan (6.2711.030)

■

Electrode stand (6.2615.130)

■

Wrench (6.1247.220)

■

Mercury collection container

■

Mercury drop catcher (6.2406.000)

1 Placing the MME pro into the electrode stand

■

Place the electrode stand into the drip pan.

■

Place the MME pro into the electrode stand.
The mercury refill opening is pointing upwards (see Figure (2-B),
Page 7
).

2 Opening the mercury reservoir

■

Use the wrench to remove the screw cap from the mercury refill
opening.

■

Use the mercury drop catcher to collect mercury adhering to the
screw cap.

■

Place the screw cap into the drip pan.

3 Pouring out the mercury

■

Pour the mercury from the mercury reservoir into the mercury col-
lection container.

■

Knock the MME pro lightly to remove remaining mercury from
the hollow spaces and pour it into the mercury collection con-
tainer.

Collect small amounts of dispersed mercury with the mercury drop
catcher.

5.2.3

Removing the capillary

Caution

In this procedure mercury is used. Always carry out the following steps
in a dry drip pan.

To remove the capillary you need the following accessories:

■

Drip pan (6.2711.030)

■

Electrode stand (6.2615.130)

■

Mercury drop catcher (6.2406.000)