5.2 Large maintenance

■■■■■■■■■■■■■■■■■■■■■■

38

■■■■■■■■

6.1246.120 Multi-Mode-Electrode pro

■

Electrode stand (6.2615.130)

■

New needle (6.1247.020)

■

Wrench (6.1247.220)

■

Scissors

1

2

3

4

Figure 12

MME pro – Inserting the needle

1 Inserting a new needle

Each needle is individually wrapped in a plastic wrapper.

Note

Do not touch the needle with bare hands!

■

Cut open the wrapper of the new needle with the scissors and
unwrap the white plastic part only. Do not yet completely remove
the wrapper.

■

Insert the needle into the hole of the membrane and press in as
far as the white marking. Now, remove the wrapper.

2 Inserting the needle holder

■

Insert the needle holder with the new needle into the MME pro.
Press in as far as it will go.

3 Removing the mounting ring

■

Unscrew the mounting ring and remove it.