background image

5.2 Large maintenance

■■■■■■■■■■■■■■■■■■■■■■

38

■■■■■■■■

6.1246.120 Multi-Mode-Electrode pro

Electrode stand (6.2615.130)

New needle (6.1247.020)

Wrench (6.1247.220)

Scissors

1

2

3

4

Figure 12

MME pro – Inserting the needle

1 Inserting a new needle

Each needle is individually wrapped in a plastic wrapper.

Note

Do not touch the needle with bare hands!

Cut open the wrapper of the new needle with the scissors and
unwrap the white plastic part only. Do not yet completely remove
the wrapper.

Insert the needle into the hole of the membrane and press in as
far as the white marking. Now, remove the wrapper.

2 Inserting the needle holder

Insert the needle holder with the new needle into the MME pro.
Press in as far as it will go.

3 Removing the mounting ring

Unscrew the mounting ring and remove it.