3.3 Properties of the RS 232 Interface

719 S Titrino

101

3.3.3 What can you do if the data transfer does not work?

Problem

Questions for remedial action

No characters can be received on

a connected printer.

- Are the instruments switched on and cables plugged in

correctly?

- Is the printer set to ”on-line”?

- Are baud rate, data bit and parity the same on both

instruments?

- Is the handshake set properly?

If everything seems to be ok, try to print a report with the key

sequence <PRINT><SMPL DATA><ENTER>. If this report

is printed out correctly, check if reports are defined in key

<DEF>.

No data transmission and the

display of the Titrino shows an

error message.

- error

42: Transmission error. Is the printer set to "on-line"?

Is the connection cable properly wired?

- error

43: Data output of the Titrino disabled for longer than 6

s by XOFF.

- error

36-39: Receive error. Are the RS settings the same on

both devices?

The received characters are

garbled.

- Are the RS settings the same on both devices?

- Has the correct printer been selected?

- Data transfer has been interrupted on the hardware side

during the printout of a curve. Re-establish connections and

switch printer off/on.

Wrong line spacing.

The printer does not emulate completely the preset mode.

Usually these problems arise with the IBM mode. Set the printer

to a different mode (e.g. Epson).

Printout of titration curve is not

ok. Other reports are printed ok.

Handshake is necessary for the printout of curves.

- Is your cable correctly wired? (The DTR of the printer has to

be connected to the CTS of the Titrino.)

- Set "HWs" for the handshake of the Titrino. Configure the

printer such that its DTR is set (possibly with DIP switches).