Metrohm 719 S Titrino User Manual

Page 33

Advertising
background image

2.7 Parameters, key <PARAM>

719 S Titrino

29


>preselections

conditioning: OFF





display drift: ON

cond.


drift corr: OFF

cond.





drift value 0.0
µl/min

cond.


req.ident: OFF

cond.





req.smpl size: OFF

cond.



activate pulse: OFF

cond.

Preselections for the sequence

Automatic conditioning of titration vessel. (ON, OFF)

If conditioning is "on", between the titrations the titration

solution is constantly maintained at the (1st) end point.

When conditioning is performed, the volume drift can

be displayed during the conditioning:

Display of drift during conditioning (ON, OFF).

Volume drift.

Type of drift correction (auto, man., OFF)

Type of drift correction: (EP – drift ∗ time)

auto: drift value at start is valid

man.: see below

OFF : no correction

Value for manual drift correction (0.0...99.9

µ

L/min)

Request of identifications after start of titration (id1,

id1&2, all, OFF)

After start, sample identifications can be requested

automatically: Only id1, id1 & id2, all three id's or no

inquiries.

Request of sample size after start of titration (value, unit,

all, OFF)

"all" the value and the unit will be requested.

Pulse output on I/O line 6 (L6, pin 1) of the remote

socket (first, all, cond., OFF)

see page 131.