2.4

Tutorial

719 S Titrino

15

Print reports

3 x <DEF>

<ENTER>

3 x <SELECT>

<;>

<ENTER>

<QUIT>

• Press <DEF> repeatedly until the display shows:

• Press <ENTER> to move to the definition of reports.

• With <SELECT> you select the individual repot

blocks. Use a ";" as delimiter between the report

blocks to print more than one block. If you wish to

print out a full result report, enter "full".

• Confirm the entry with <ENTER> and quit the inquir-

ies with <QUIT>.

Now print your reports by pressing <PRINT><REPORTS><ENTER>. Your printout will

look like the following:

Identification of the report type (fr=full report)

Device type, device number and program version

Initial pH value, mode and method name

Volume and pH value of EP1

Calculated result

The development of your method is now complete. Before we store it in the method

memory, you should check it again. Prepare a fresh sample and restart the titration with

<START>.

If everything appears to be all right, you can now store the method in the method mem-

ory.

def

>report

report:

report:curve;full

'fr

719 S Titrino 18200 719.0021

date 2002-03-13 time 17:34 1

pH(init) 2.40 SET pH ********

smpl size 2 ml

EP1 2.032 ml 8.14

RS1 3.71 g/l

---------------