Metrohm 719 S Titrino User Manual

Page 9

Advertising
background image

1.2 Controls and parts

719 S Titrino

5

6

RS232

interface

for the connection of printer, balance or a computer

7

Remote

lines

(input/output)

for the connection of the Remote Box, Sample Changers, robots etc.

8

Connection of electrodes and temperature sensor
• 2 high-impedance measuring inputs for pH and U measurements (Ind I/ Ind II).

They can either be used separately or for differential potentiometry, see page

124.

Important: If you work with both measuring inputs in the same vessel, the same

reference electrode must be used.

• 1 measuring input for polarized electrodes (Pol).
• 1 measuring input for PT100 or Pt1000 temperature sensor.

9

Connection for separate keypad

10

Analog output for the connection of a recorder

11

Connection for power cable

With power supplies where the voltage is subject to severe HF disturbances, the Ti-

trino should be operated via an additional power filter, e.g. Metrohm 615 model.

12

Mains

switch

13

Earthing

socket

14

Connection for stirrer

728 Magnetic Stirrer, 802 Rod Stirrer, 703 or 727 Ti Stand
Supply voltage: 9 VDC (I ≤ 200 mA)

15

Display of the set mains voltage

Before switching on for the first time, check that the set mains voltage matches the

voltage of your power supply. If this is not the case, disconnect mains cable and

change voltage.

16

Rating

plate

with fabrication, series and instrument number