background image

2.4. Tutorial

719 S Titrino

14

Entry of the sample size

4 x <SMPL DATA>

<2>

<ENTER>

<SELECT>

<ENTER>

• Press <SMPL DATA> repeatedly until "sample size"

appears in the display.

• Enter 2.

• Use <SELECT> to select the unit "mL" and confirm

the new value with <ENTER>.

The result is now calculated and can be displayed in place of the end point. If your

method already includes a formula at the end of the titration, the calculated result is dis-

played directly after the titration. As we have entered the formula later, we now have to

select the result display:

Display of the result

<SELECT>

<ENTER>

• Press <SELECT> repeatedly until ">display results"

appears in the display.


• Press <ENTER> to move to the result display.

If you have a printer connected, you probably wish to have the curve and a result report

printed out automatically at the end of a titration.

To install a printer, see page 119.

smpl data

smpl size

1.0 g

smpl data

smpl unit:

g

SET pH

********

>display results

SET pH

********

RS1 3.71

g/l