3 Installation

■■■■■■■■

13

The metal pin on the sample cup holder must go into the indentation
on the sampling window.

2

■

In the software, start Wavelength Certification. Follow the
instructions in the software.

■

When the software prompt appears, insert the wavelength stan-
dard (WSR12007) into the opening of the holder. The label of the
wavelength standard must be visible.

■

Close the lid.

■

Wait until Wavelength Certification is completed.

■

Remove the wavelength standard from the holder.