Metrohm NIRS DS2500 Analyzer User Manual

Page 22

Advertising
background image

3.6 Initial start-up

■■■■■■■■■■■■■■■■■■■■■■

14

■■■■■■■■

3

■

In the software, start Reference Standardization. Follow the
instructions in the software.

■

When the software prompt appears, insert the reflection standard
(R8012007) into the opening of the holder. The label of the
reflection standard must be visible.

■

Close the lid.

■

Wait until Reference Standardization is completed.

■

Remove the reflection standard from the holder.

Regulated industries

If you put the NIRS DS2500 Analyzer into operation in a regulated indus-
try, then additional tests are required after the procedures described in the
chapter "Non-regulated industries" have been performed.

In regulated industries, the Vision Pharma software version and the NIRS
reflection standard, 7 pcs. (6.7450.010) are required for the instrument to
be put into operation.

Information on which additional tests are required for your specific regu-
lated industry can be found in the applicable standards.