3 inserting the sample cup for liquid samples, Inserting the sample cup for liquid samples – Metrohm NIRS DS2500 Analyzer User Manual

Page 25

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

■■■■■■■■

17

3.7.3

Inserting the sample cup for liquid samples

Inserting the sample cup for liquid samples

Required accessories

■

NIRS liquid sample kit transflection (6.7400.010)

■

NIRS DS2500 holder for sample cup (6.7430.040)

1 Clean the sample cup and the sampling window with a lens cleaner

cloth.

2

■

Fill liquid sample to a height of approx. 1 cm into the sample cup.

■

Place the gold diffuse reflector into the liquid. Avoid air pockets
while doing so.

3

■

Place the sample cup holder onto the sampling window.
The metal pin on the holder must go into the indentation on the
sampling window.

■

Place the sample cup into the round opening of the holder.